5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
526
■■■■■■■■
viva 1.1 (for process analysis)
Moves the selected line upward (modifies sequence).
Moves the selected line downward (modifies sequence).
[New]
Opens the Fixed point evaluation # dialog window for entering a new
fixed endpoint (see Chapter 5.5.2.7.4.6, page 529).
[Properties]
Opens the Fixed point evaluation # dialog window to edit the fixed
point selected in the table (see Chapter 5.5.2.7.4.6, page 529).
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this check box is activated, then additional parameters will be displayed
for the minimum measured value from the measuring point list.
Threshold value
The evaluation of the minimum begins as soon as the absolute value of
the slope of the curve exceeds the set threshold value.
Input range
0.1 - 1,000.0 mAU/s
Default value
20.0 mAU/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this check box is activated, then additional parameters will be displayed
for the maximum measured value from the measuring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Input range
0.1 - 1,000.0 mAU/s
Default value
20.0 mAU/s