2 prep - properties – Metrohm viva 1.1 (ProLab) User Manual
Page 480
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
468
■■■■■■■■
viva 1.1 (for process analysis)
Parameters
The parameters for the PREP command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the PREP command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
CYL
Cylinder volume of the exchange or dosing unit used
for the command
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started
5.5.2.4.7.2
PREP - Properties
Dialog window: Method
▶ PREP ▶ Properties... ▶ PREP - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined