5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
610
■■■■■■■■
viva 1.1 (for process analysis)
Identification
Description
ELT
Variable for electrode test:
0 = Electrode test has not been performed
1 = Electrode test has been performed once
2 = Electrode test has been performed twice
3 = Electrode test has been performed three times
FIN
Command status:
1 = Command has been called and ended at
least once
0 = Command has been called and never
ended
invalid (variable not available) = Command has
never been called
HR
Highest current measuring range used:
1 = 224 mA | 3 = 20 mA | 5 = 2 mA | 7 = 200 µA |
9 = 20 µA | 11 = 2 µA | 13 = 200 nA | 15 = 20 nA |
17 = 2 nA | 19 = 200 pA
LR
Lowest current measuring range used:
19 = 200 pA | 17 = 2 nA | 15 = 20 nA | 13 = 200 nA
| 11 = 2 µA | 9 = 20 µA | 7 = 200 µA | 5 = 2 mA |
3 = 20 mA | 1 = 224 mA
NMP
Number of measuring points in measuring point list
RPM
Stirring rate in min
–1
STY
Type of stop with which the command was stopped:
1 = finished normally, 0 = stopped manually
with [Quit] or [Stop]
Result variables if called by CALL VMS
The following result variables are generated in the method run when a
CVS command is called by CALL VMS and can be used in formulas:
Result variables per substance, VMS and replication:
Designation 'RS.'Command name'.'Substance
name'.VMS.REP{y}.Variable identification'
Identification
Description
SUBST
Substance name