3 transfer - transfer command – Metrohm viva 1.1 (ProLab) User Manual
Page 519
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
507
[Properties]
Opens the Transfer command dialog window for editing the transfer
command selected in the table (see Chapter 5.5.2.6.6.3, page 507).
[Delete]
Deletes the selected transfer command.
Timeout
on | off (Default value: off)
If this check box is activated, then a maximum waiting time can be
entered. If not all transfer commands have finished after this time period,
the command is canceled and the next command is started.
Input range
0 - 9,999.9
Default value
1
Selection
min | s
Default value
s
5.5.2.6.6.3
TRANSFER - Transfer command
Dialog window: Method
▶ TRANSFER ▶ Properties... ▶ Transfer command
Name
Freely definable name for the transfer command.
Entry
50 characters
Action
Selection
Read | Write | Wait for | Poll
Default value
Read
Read
The character string (optional) defined in Command is sent to the
instrument and the answer is expected and evaluated within the Time-
out defined in the instrument settings.
Write
The character string defined in Command is sent to the instrument.
Wait for
The character string (optional) defined in Command is sent to the
instrument and a waiting period ensues that is as long as it takes for a
received character string to correspond to the regular expression in
Comparison. If this does not take place within the Timeout defined
in the command, the command will be canceled.