4 properties - special beaker, Special beakers – Metrohm viva 1.1 (ProLab) User Manual
Page 848
6.12 Rack data subwindow
■■■■■■■■■■■■■■■■■■■■■■
836
■■■■■■■■
viva 1.1 (for process analysis)
Input range
0 - 235 mm
Tower 2
Defines the lift positions for tower 2. These apply for all rack positions
except those that are defined as Special beakers.
Work position
Work position for lift 2.
Input range
0 - 235 mm
Rinse position
Rinse position for lift 2.
Input range
0 - 235 mm
Shift position
Shift position for Lift 2. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then it will shift at the current
lift position. This means that the shift position must be selected so that a
safe movement across the entire rack is possible at any time.
Input range
0 - 235 mm
Special position
Special position for lift 2.
Input range
0 - 235 mm
6.12.3.4
Properties - Special beaker
Dialog window: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack
data
▶ Special beakers
The table shows the data of all special beakers of the attached rack in tab-
ular form. Any position on the rack can be assigned to each special
beaker. The table cannot be edited directly.
NOTE
Special beakers should preferably be set at high rack positions in order
to be able to begin sample series at rack position 1. Rack positions
defined as special beakers are no longer available as sample positions. A
specific work position can be defined for each special beaker at tower 1
and tower 2 (if present). The shift, rinse and special positions of the cor-
responding tower are applied from the general rack positions.