5 Method
viva 1.1 (for process analysis)
■■■■■■■■
549
Input range
0.1 - 600.0 s (Increment: 0.1)
Default value
1.0 s
Signal drift
The measurement is canceled if the value has fallen below the signal drift
defined here.
Input range
0.1 - 999.0 °C/min
Default value
0.1 °C/min
Min. waiting time
The signal drift is only checked after the minimum waiting time defined
here has been exceeded.
Input range
0 - 6,500 s
Default value
0 s
Max. waiting time
The measurement is canceled as soon as the maximum waiting time
defined here has been exceeded. If the waiting time has not been newly
entered, then a waiting time that is suitable for the drift will be calculated
automatically according to the following formula:
Input range
0 - 6,500 s
Default value
457 s
Stop measured value
The measurement is canceled when the specified measured value since
the start of the measurement is reached.
The measurement is also canceled if the Min. waiting time has not yet
been exceeded.
It will also be canceled if the value has not yet fallen below the Signal
drift.
Input range
0.0 - 100.0 °C
Selection
off
Default value
off
off
No stop.
Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed: