5 Method
viva 1.1 (for process analysis)
■■■■■■■■
637
[New]
Opens the Stripping dialog window, in which the stripping potential and
the stripping time of a potential step can be entered ((see Chapter
5.5.2.11.3.7, page 639)).
[Properties]
Opens the Stripping dialog window, in which the stripping potential and
the stripping time of a potential step can be edited (see Chapter
5.5.2.11.3.7, page 639).
[Delete]
The selected potential step is deleted.
Cleaning potential
Potential that is applied to the electrodes during the Cleaning time.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Default value
1.625 V
Cleaning time
Waiting time during which the Cleaning potential is applied to the elec-
trodes.
Input range
0.0 - 50.0 s (Increment: 0.1)
Default value
5.0 s
Measuring interval
Measuring interval for data acquisition during plating and stripping.
Selection
1.28 ms | 2.56 ms | 5.12 ms | 10.24 ms | 20.48
ms | 40.96 ms | 81.92 ms
Default value
5.12 ms
Preparation cycles
Number of cycles without current measurement that are run through
before the measuring cycles.
Input range
0 - 200 (Increment: 1)
Default value
2
Measuring cycles
Number of cycles with current measurement.
Input range
1 - 50
Default value
2
Sweep duration
Shows the duration of the sweep in s.