5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
444
■■■■■■■■
viva 1.1 (for process analysis)
Selection
814 USB Sample Processor | 815 Robotic USB
Sample Processor XL | 858.0010 Professional
Sample Processor | 858.0020 Professional
Sample Processor | 858.0030 Professional
Sample Processor | 919.0020 IC Autosampler
plus
Default value
858.0020 Professional Sample Processor
Rack test
on | off (Default value: off)
If this check box is activated, then a check will be made to determine
whether the correct rack is in place. If this check box is not activated, then
the only check made will be to determine whether any rack has been
mounted.
Test rack
Selection of the rack name for the rack that needs to be mounted. You
can thus ensure that the method can be executed only with this rack. If a
different rack is detected when the command is executed, then a message
will be displayed and the exit track (if available) will be called.
Selection
all of the sample racks present in the configu-
ration
5.5.2.3.8
HEATER
5.5.2.3.8.1
HEATER - Overview
Dialog window: Method
▶ HEATER ▶ Properties... ▶ HEATER - 'Command name'
Command for the control of the heater.
Instruments
This command can be executed with the following instruments:
Photometer: 089
Appearance
The command has the following appearance:
Parameters
The parameters for the HEATER command are configured in the follow-
ing dialog window:
■