5 Method
viva 1.1 (for process analysis)
■■■■■■■■
587
Live display
on | off (Default value: on)
If this check box is activated, then the track will be displayed in the live
display.
Return immediately
on | off (Default value: off)
If this check box is activated, then the program does not wait for this track
to be completed, but rather this track responds immediately to the track
with the call command. From now on, both tracks will be running simulta-
neously.
Command variables
The following command variables are generated by the command VA
TRACK in the method run and can be used in formulas under the desig-
nation 'Command name.Variable identification':
Identifi-
cation
Description
BSY
Command status:
1 = BUSY, HOLD or ERROR
0 = READY
invalid (variable not available) = Command has never
been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
invalid (variable not available) = Command has never
been started
5.5.2.9.5
SERIES START
Dialog window: Method
▶ SERIES START ▶ Properties... ▶ SERIES START - 'Com-
mand name'
Start command for the series start track.
Appearance
The command has the following appearance: