5 Method
viva 1.1 (for process analysis)
■■■■■■■■
547
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
894 Professional CVS
Default value
894 Professional CVS
Sensor
Sensor
Selection of a sensor of the type Temperature sensor from the list of
sensors available in the sensor table.
Selection
'Sensor name' | Temperature sensor
Default value
Temperature sensor
Stirrer
Stirring rate
Number of revolutions per minute of the stirrer.
Input range
0 - 3,000 min
⁻¹ (Increment: 100)
Default value
2,500 min
⁻¹
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.