Metrohm viva 1.1 (ProLab) User Manual
Page 502
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
490
■■■■■■■■
viva 1.1 (for process analysis)
Signal drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.
Input range
0.1 - 999,000 mAU/min
Default value
1.0 mAU/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the minimum waiting time is elapsing.
Input range
0 - 999,999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:
Input range
0 - 999,999 s
Default value
154 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Input range
0.1 - 999,999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 4,000 mAU
Selection
off
Default value
off
off
No stop.