5 Method
viva 1.1 (for process analysis)
■■■■■■■■
645
Reference electrode
Selection of an electrode of the Reference electrode type from the elec-
trodes available in the electrode table.
Selection
Reference electrode | 'Electrode name'
Default value
Reference electrode
Reference electrode
Standard electrode of the Reference electrode type, which cannot
be deleted.
'Electrode name'
Electrode of the Reference electrode type from the electrode table.
Auxiliary electrode
Selection of an electrode of the Auxiliary electrode type from the elec-
trodes available in the electrode table.
Selection
Auxiliary electrode | 'Electrode name'
Default value
Auxiliary electrode
Auxiliary electrode
Standard electrode of the Auxiliary electrode type, which cannot be
deleted.
'Electrode name'
Electrode of the Auxiliary electrode type from the electrode table.
Electrode check
The electrode test for the reference and auxiliary electrodes can be acti-
vated and deactivated only in the START command (see "Electrode test
for VA measuring commands", page 579).
Stirrer
Stirring rate
Number of revolutions per minute of the stirrer.
Input range
0 - 3,000 min
⁻¹ (Increment: 100)
Default value
2,500 min
⁻¹
Hydrodynamic measurement
on | off (Default value: on)
If this check box is activated, then stirring will continue during the sweep.
Only editable for Sensor type = RDE, otherwise disabled.