Metrohm tiamo 2.3 (process analysis) User Manual
Page 975
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
959
Instruments
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 851, 852, 857, 888, 890, 901, 904,
905, 906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command MEAS Ipol are configured on the fol-
lowing 4 tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Evaluations
Defining further methods for evaluating measurement curves.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS Ipol and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BP{x}.MEA
Measured value for the break point x (1…9) in the unit
of the measured value
.BP{x}.TEM
Temperature for the break point x (1…9) in °C
.BP{x}.TIM
Time for the break point x (1…9) in s