Metrohm tiamo 2.3 (process analysis) User Manual
Page 646
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
630
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Measured value acceptance
Note
Will be displayed only if the option User is selected in the selection list
Titration rate.
Measured value drift
Drift for the measured value acceptance during the titration.
Titrando, 855
Range
0.1 ... 999.0 mV/min
Default value
20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Titrino
Range
0.5 ... 999.0 mV/min
Default value
20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Titrando, 855
Range
0 (slow, optimal, fast) ... 999999 s
Default value
0 (slow, optimal, fast) s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting