5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1258
■■■■■■■■
tiamo 2.3 (for Process Analysis)
5.6.9.4.3
SEND - Send event message
Dialog window: Method
▶ SEND ▶ Properties... ▶ [New]/ [Properties] ▶ Send event
message
Receiver
Selection of the type of receiver.
Selection
System | Command
Default value
System
Selection
Selection of the receiver address. All of the command names already
defined in the method are available for Receiver = Command. This field
is empty and cannot be edited for Receiver = System.
Selection
'Command name' | 'empty'
Default value
'empty'
Event message
Selection of the event message to be sent. The following event messages
can be selected depending on the type of receiver:
System
Selection
Quit | Hold | Stop
Default value
Quit
Command
Selection
Quit | Hold | Continue | Start Titration
Default value
Quit
Receiver
Selec-
tion
Event
mes-
sage
Significance
System
-
Hold
Hold determination (all tracks).
This corresponds to the [Hold]
button in the dialog window Run
in the program part Workplace.
System
-
Quit
Cancel determination (Series con-
tinues).
System
-
Stop
Stop determination and series.
This corresponds to the [Stop]
button in the dialog window Run
in the program part Workplace.