5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1228
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Init.temp.
With Init.temp. the cuvette heating is switched on at the same time
as the device and the cuvette is heated up to this temperature. The ini-
tial temperature is entered and altered in tiamo.
off
Switch off the heater.
5.6.7.9
FLOW
5.6.7.9.1
FLOW - Overview
Dialog window: Method
▶ FLOW ▶ Properties... ▶ FLOW - 'Command name'
Command for controlling the gas flow.
Devices
This command can be executed with the following devices:
Sample Processor: 774, 874
Appearance
The command has the following appearance:
Parameters
The parameters for the command FLOW are configured in the following
dialog window:
■
Command variables
The following command variables are generated in the method run by the
command FLOW and can be used in formulas under the designation
'Command name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended