Metrohm tiamo 2.3 (process analysis) User Manual
Page 1661
■■■■■■■■■■■■■■■■■■■■■■
7 Instruments
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1645
Default value
Display of the default value the outputs takes directly after starting and
exiting tiamo. The default value is also set after an error or when cancel-
ing the method.
Note
Behavior at Emergency stop:
■
All digital outputs except the ones of the Type = Alarm signal are
set to the default value when an emergency stop takes place.
■
A digital output of the Type = Alarm signal is set to the non-
default value when an emergency stop takes place.
■
Digital outputs assigned to a digital input of the Type = Level
display or Leak detector do not change their status when an
emergency stop takes place.
Selection
active | inactive
Default value
inactive
Default values for commands
Parameter 1
The meaning of Parameter 1 depends on the Type of the digital output
that has been set.
When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.
Range
-214783647 ... 214783647
Default value
100
Parameter 2
The meaning of Parameter 2 depends on the Type of the digital output
that has been set.
For Type = Stirrer:
■
0: Duration of the switch-on pulse fixed at 500 ms.
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in
10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on
225 ms, switched off 75 ms