6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1437
Name
Name of the colorimetric sensor.
Calibration date
Display of date and time of the last calibration entered after each auto-
matic calibration or manual entry.
Wavelength
Wavelength at which the calibration of the instrument has been carried
out.
Coefficient of determination
Display of the coefficient of determination r
2
, which is calculated on the
basis of the calibration function and the confidence interval and which is
between 0 und 1. 5 decimal places are displayed.
Confidence interval
Display of the confidence interval within which the measured values for
the determination of the calibration curve must lie.
c0
Calibration coefficient of zero order.
c1
Calibration coefficient first-order.
c2
Calibration coefficient second-order.
c3
Calibration coefficient third-order.
Additional columns from the properties of the colorimetric sensors can be
displayed with the menu item Edit
▶ Column display....
Note
Lines that contain red entries will also show the line number with a red
background.
Table view
The table of the colorimetric sensors cannot be edited directly. With a click
on the column title the table can be sorted according to the selected col-