795
5.6.3.5.2.5
KFT Ipol - Titration parameters
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Titration direction
Selection of the titration direction.
Selection
+ | - | auto
Default value
auto
+
positive measured value modification
-
negative measured value modification
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.
Range
0 ... 999999 s
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C