Metrohm tiamo 2.3 (process analysis) User Manual
Page 680
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
664
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.5 ... 10.0 µA/mL
Default value
5.0 µA/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.5 ... 10.0 µA/mL
Default value
5.0 µA/mL
Break point evaluation
Note
This evaluation method is possible only with Titrando and 855.
Break-point evaluation
on | off (Default value: off)
A break-point evaluation is used to determine sharp changes of direction
in the titration curve.
EP criterion
Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value