5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1104
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
5.6.5.5.4
CAL Spec - Calibration wavelengths
Tab: Method
▶ CAL Spec ▶ Properties... ▶ Calibration wavelengths
Command name
Name of the command.
Input
25 characters
The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.
Tolerance
Tolerance for the preset wavelength when assigning the highest peak
maximum found during the calibration measurement to the reference
wavelength. If a peak is outside the range Wavelength +/- Tolerance,
the calibration will be canceled. The existing calibration in the configura-
tion remains in effect.
Range
0.1 ... 1100.0 nm
Default value
0.5 nm
[New]
Open the dialog window Calibration wavelength # to enter wave-
lengths and tolerances (see Chapter 5.6.5.5.5, page 1104).
[Properties]
Open the dialog window Calibration wavelength # to edit the values
selected in the table (see Chapter 5.6.5.5.5, page 1104).
[Delete]
Delete the selected line.
5.6.5.5.5
CAL Spec - Calibration wavelength #
Tab: Method
▶ CAL Spec ▶ Properties... ▶ Calibration wavelengths ▶ [New]/
[Properties]
▶ Calibration wavelength #
Wavelength
Range
200.0 ... 1100.0 nm
Default value
'empty' nm