5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
552
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop measured value
Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.
Range
-200.0 ... 200.0 µA
Selection
off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Range
1 ... 9
Default value
9
Selection
off
Volume after EP
When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.
Titrando, 855
Range
0.01000 ... 9999.99 mL
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.