5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
868
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command. This parameter is displayed only for 751, 785, 798
and 799.
5.6.3.8.6.3
STAT pH - Start conditions
Tab: Method
▶ STAT pH ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titra-
tion is started.
Initial measured value
Note
Only displayed for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
The measured value is only accepted if the drift is less than the value
entered here.
Range
0.1 ... 999.0 mV/min
Selection
off
Default value
off