3 met cond - start conditions, Start conditions – Metrohm tiamo 2.3 (process analysis) User Manual
Page 694
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
678
■■■■■■■■
tiamo 2.3 (for Process Analysis)
5.6.3.3.6.3
MET Cond - Start conditions
Tab: Method
▶ MET Cond ▶ Properties… ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the con-
ductivity measurement is started.
Initial measured value
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Signal drift
The measured value is only accepted if the drift is less than the value
entered here.
Range
0.001 ... 999.0 (mS/cm)/min
Selection
off
Default value
off
off
The measured values will be accepted when the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s