6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1429
Message
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the e-mail address defined under [E-
mail...] if this check box is activated. The message is sent in text format.
Can be edited only for Monitor ELT validity = on, otherwise inactive.
[E-mail...]
Opens the Send e-mail window (see Chapter 2.6.1, page 92).
Active only if Message per e-mail = on, otherwise inactive.
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message defined above.
Can be edited only for Monitor ELT validity = on, otherwise inactive.
Action
If during monitoring it is found that the validity period has expired then
one of the following actions will be triggered automatically at start test.
Can be edited only for Monitor ELT validity = on, otherwise inactive.
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the
determination.
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].