Metrohm tiamo 2.3 (process analysis) User Manual
Page 1649
■■■■■■■■■■■■■■■■■■■■■■
7 Instruments
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1633
Input parameters
Tolerance range
for control task = digestion unit
Range
0 ... Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
1.0 'Mapping unit of the input or of the cal-
culated value'
for control task = free controller
Range
0 ... Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
1 'Mapping unit of the input or of the calcula-
ted value'
Working area
for control task = digestion unit
Range
0 ... Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
40.0 'Mapping unit of the input or of the cal-
culated value'
for control task = free controller
Range
0 ... Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
10 'Mapping unit of the input or of the calcu-
lated value'
Output parameters
PMW cycle duration
for control task = digestion unit
Range
0 ... 9999.0 s
Default value
1.0 s