3 stirring - continuous operation, 3 remote functions, Remote functions 0 – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1676
8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1660
■■■■■■■■
tiamo 2.3 (for Process Analysis)
8.4.2.3
Stirring - Continuous operation
Program part: Manual control
Here stirrers can be switched on for a defined time. This tab only appears
if a single stirrer is selected.
Stirring rate
Selection of the stirring rate. This parameter can also be modified live.
all instruments except for photometer
Range
–15 ... 15
Default value
8
Photometer only
Range
1 ... 15
Default value
3
Stirring time
Entry of the time during which stirring is to be carried out. If this parame-
ter is modified after the start of the stirrer, it is not valid until the next stir-
ring procedure.
Range
1 ... 999999 s
Default value
60 s
[Start]
Start continuous operation for the selected stirrer. In the status display the
remaining time is shown. The stirrer will be switched off automatically
after the stirring time has elapsed
[Stop]
Stop stirring for the selected stirrer.
8.4.3
Remote functions
Program part: Manual Control
If in the subwindow for the device selection, a remote box or a remote
interface is selected, then the functions and parameters belonging to
these elements are displayed in the subwindow Functions/Parameters.
Devices
The remote functions can be carried out with the following devices:
Titrino: 702*, 716*, 718*, 719*, 720*, 721*, 736*, 751, 758, 784, 785,
794*, 795, 798, 799 (* devices with only 3 output lines)
Titrando: 808, 809, 835, 836, 841, 842, 857, 888, 890, 901, 902, 904,
905, 906, 907