Metrohm tiamo 2.3 (process analysis) User Manual
Page 1420
6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1404
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Order number
Order number of dosing unit. With intelligent dosing units the order num-
ber is read off automatically and cannot be edited.
Input
24 characters
Default value
'empty'
Serial number
Serial number of dosing unit. With intelligent dosing units the serial num-
ber is read off automatically and cannot be edited.
Input
10 characters
Default value
'empty'
Cylinder volume
Cylinder volume of dosing unit. With intelligent dosing units the cylinder
volume is read off automatically and cannot be edited. If in the method
you have selected a solution in a titration or dosing command then the
cylinder volume will be checked in the run.
Selection
2 mL | 5 mL | 10 mL | 20 mL | 50 mL
Default value
20 mL
Cylinder serial number
Serial number of cylinder. It is printed on new cylinders. With intelligent
dosing units it is read off automatically. It can be altered at any time, e.g.
if the cylinder has to be replaced.
Input
8 characters
Default value
'empty'
Parameters for preparation
Configures the parameters to be used in the commands PREP and
EMPTY.
Dosing port Prep/Empty
Dosing port via which the cylinder contents are to be ejected during prep-
aration and emptying.
Selection
Dosing port 1 | Dosing port 2 | Fill port | Spe-
cial port
Default value
Dosing port 1
Dosing rate Dosing port 1
Speed at which dosing is to be carried out at Dosing port 1. The maxi-
mum dosing rate depends on the cylinder volume of the exchange unit