5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
787
Input
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Titrando (without 890)
Selection
1 | 2
Default value
1
890
Selection
1
Default value
1
Sensor
Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.
Titrando, 855
Range
–125.0 ... 125.0 µA (Increment: 0.5)
Default value
5.0 µA
Titrino
Range
–127 ... 127 µA (Increment: 1)
Default value
5 µA
Electrode check
on | off (Default value: off)
If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.