5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1242
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Save result as common variable
on | off (Default value: off)
The result will be saved under the selected name as common variable if
this option is activated (see Chapter 6.9.1, page 1447). In all cases, only
the single value for this result will be saved as a common variable, even if
the statistics are enabled for this result.
Common variable
The result is saved as new value for this common variable.
Selection
List of all the common variables defined
Note
As long as no common variable has been defined, the selection box will
be empty. The creation of common variables is described in the config-
uration (see Chapter 6.9.1, page 1447).
Note
If common variables are used in a CALC command, then they will only
be requested at the start of the determination. This means that it is not
possible to write Common variables during a determination and then
use the modified value in a different formula. For this you should use
Method variables (see Chapter 2.4.3.2, page 33).
Save result as titer
on | off (Default value: off)
If this option is enabled, then the result will be saved as a titer under the
selected name. If statistics is activated for this result, then the current
mean for this result will automatically be saved as titer.
Solution name
Name of the solution for which the result is to be saved as titer.
Selection