5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
834
■■■■■■■■
tiamo 2.3 (for Process Analysis)
756, 831
Range
0.3 ... 999.9 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
851, 852
Range
0.3 ... 999.0 µg/min
Default value
0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min
Selection
minimum
Stop criterion
Stop criterion
Switches off titration when the endpoint has been reached and the stop
criterion has been met.
756, 831
Selection
Drift | rel. drift
Default value
rel. drift
851, 852
Selection
Drift | rel. drift | Time | off
Default value
rel. drift
Stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.
Range
1 ... 999 µg/min
Default value
5 µg/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Range
0 ... 999 µg/min
Default value
5 µg/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time.
Range
0 ... 999 s
Default value
10 s