5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
941
Stop measured value pH
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-20.000 ... 20.000
Selection
off
Default value
off
off
No stop.
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Titrando, 855, 867
Range
-20.0 ... 150 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500 °C
Default value
25.0 °C
5.6.4.3.4
MEAS pH - Evaluations
Tab: Method
▶ MEAS pH ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following methods for evaluation of measurement curves can be acti-
vated and defined on this tab.
Fixed endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fix measured quantity (Meas-
ured value or Time) for the fix endpoint from the measuring point list.