5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
901
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stop when the entered time has elapsed after the start conditions have
been fulfilled.
Range
0 ... 999999 s
Selection
off
Default value
off
since
Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after
the following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i. e. the
stop time is reset to zero again after each dosing step.
Stop rate
Rate at which the dosing cylinder is to be filled after titration. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off