5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1101
Reference temperature
Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.
Range
-20.0 ... 150.0°C
Default value
20.0
Conductivity
Input of the conductivity of the standard at the reference temperature.
Range
0.00000 ... 2000.00 mS/cm
Default value
1.0 mS/cm
Temperature coefficient
Temperature coefficient of the standard with the reference temperature.
Range
0.00 ... 9.99 %/°C
Default value
1.00 %/°C
5.6.5.5
CAL Spec
5.6.5.5.1
CAL Spec - Overview
Dialog window: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command
name'
Command for wavelength calibration.
Instruments
This command can be executed with the following devices:
Spectrometer: Avantes
Appearance
The command has the following appearance:
Parameters
The parameters for the command CAL Cond are configured on the fol-
lowing 3 tabs:
■
General/Hardware
Parameters for device.
■
Measuring parameters
Parameters for setting the measurement procedure.