Metrohm tiamo 2.3 (process analysis) User Manual
Page 899
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
883
[Properties]
Open the dialog window Evaluation window # in which the parameters
for the new time slot can be edited (see Chapter 5.6.3.8.3, page 858).
[Delete]
Delete the time slot selected in the table.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint.
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.4.3.10, page 714).
[Delete]
Delete the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 20.0 pH/mL
Default value
1.0 pH/mL