Metrohm tiamo 2.3 (process analysis) User Manual
Page 1087
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1071
Titrando: 808, 809, 835, 836, 857, 904, 905, 906, 907
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command STDADD dos are configured on the
following 3 tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Standard addition
Parameters for standard addition.
■
Measuring parameters
Parameters for setting the measurement procedure.
Standard addition curve
The standard addition curve with its associated evaluation data for a
determination is shown in the program part Database in the dialog win-
dow Calibration curve, which is opened with the menu item Determi-
nations
▶ Show calibration curve... or the symbol .
Command variables
The following command variables are generated in the method run by the
command STDADD dos and can be used in formulas under the designa-
tion 'Command name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used for
the command
.DBL
Total duration for the processing of the command in s