3 solution properties, 1 editing solution properties, 2 properties - solution – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1408: Solution properties 2
6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1392
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Landscape
Print the solution table in landscape format.
[OK]
The solution table is shown in the required format as a PDF file and can be
opened directly with Acrobat Reader; it can then be printed out and/or
saved.
6.6.3
Solution properties
6.6.3.1
Editing solution properties
Dialog window: Configuration
▶ Titrants/Solutions ▶ [Edit] ▶ Properties… ▶ Solu-
tion - 'Name'
The parameters for the selected solution are defined on the following
tabs:
■
Solution
Properties of the solution and solution monitoring.
■
Titer
Information about the titer value and titer determination.
■
Titer history
Display of the last 10 titer values.
■
Exchange unit
Properties of the exchange unit containing the solution.
■
Dosing unit
Properties of the dosing unit containing the solution.
■
GLP
Properties of GLP test and GLP monitoring.
6.6.3.2
Properties - Solution
Tab: Configuration
▶ Titrants/Solutions ▶ [Edit] ▶ Properties... ▶ Solution - 'Na-
me'
▶ Solution
Solution name
Name of the solution (can be entered or selected).
Input
24 characters
Selection
Selection from predefined names
Concentration (value)
Concentration value of the solution. This value for the concentration of a
solution to be used in the titration commands DET, MET, SET and KFT is
available as a variable 'Command name.CONC' for calculations.
Range
0.0 ... 1E+13
Default value
1.000