8 set ipol - additional evaluations, Additional evaluations – Metrohm tiamo 2.3 (process analysis) User Manual
Page 771
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
755
Note
Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the
titration will be ignored.
Show measured value during conditioning
on | off (Default value: off)
If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.
5.6.3.4.5.8
SET Ipol - Additional evaluations
Tab: Method
▶ SET Ipol ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.4.5.10, page 758).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.4.5.10, page 758).