5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1291
Parameter 2
The meaning of Parameter 2 depends on the Type of the digital output
that has been set.
For Type = Stirrer:
■
0: Duration of the switch-on pulse fixed at 500 ms.
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in
10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on
225 ms, switched off 75 ms
Range
-214783647 ... 214783647
Default value
0
5.6.11.3
DIGITAL IN
5.6.11.3.1
DIGITAL IN - Overview
Dialog window: Method
▶ DIGITAL IN ▶ Properties... ▶ DIGITAL IN - 'Command
name'
Devices
This command can be executed with the IO Controller and with the Step-
ping Motor Controller.
Appearance
The command has the following appearance:
Mode of operation
The statuses of all digital inputs will be entered and saved in command
variables at the time of the command (syntax Command name.Variable
designation)
Parameters
The parameters for the command DIGITAL IN are configured on the fol-
lowing 2 tabs:
■
General
Parameters for device.
■
Stop criteria
Parameters for defining the stop criteria.