5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1016
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.4.10.3
MEAS Cond - Measuring parameters
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Type of measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is selec-
ted. The measurement is canceled as soon as the defined Signal drift
or the Stop measured value is reached or a Waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled. This option is not available for the 712 Conductometer.
Measurement without drift control
The measurement is carried out without drift control if this option is
selected. The measurement continues until one of the stop criterions
Measuring time or Stop measured value is met. The correspond-
ing parameters are not visible if this option is disabled.