3 meas spec - measuring parameters – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1052
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1036
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Avantes Spectrometer
Default value
Avantes Spectrometer
5.6.4.13.3
MEAS Spec - Measuring parameters
Tab: Method
▶ MEAS Spec ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Start wavelength
Lower limit of the spectrum.
Range
100.0 ... 2000.0 nm
Default value
400.0 nm
End wavelength
Upper limit of the spectrum.
Range
100.0 ... 2000.0 nm
Default value
1000.0 nm
Integration time
Integration time for spectrum recording.
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.