11 det upol - fixed endpoint evaluation, 3 met, 1 met - overview – Metrohm tiamo 2.3 (process analysis) User Manual
Page 579
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
563
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.5.11
DET Upol - Fixed endpoint evaluation
Dialog window: Method
▶ DET Upol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation # ▶ [New]/[Property] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Range
-200.0 ... 200.0 µA
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.3
MET
5.6.3.3.1
MET - Overview
Menu item: Method
▶ Insert ▶ New command...
Command for Monotonic equivalence point titrations.