5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1112
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.5.7.3
CAL MEAS Conc - Measuring parameters
Tab: Method
▶ CAL MEAS Conc ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The measuring parameters for the calibration are defined on this tab.
Measurement with drift control
The ISE calibration is carried out drift-controlled. The measurement is stop-
ped as soon as the Signal drift is reached or a defined waiting time has
elapsed.
Measured value drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.
Range
0.1 ... 999.0 mV/min
Default value