Metrohm tiamo 2.3 (process analysis) User Manual
Page 688
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
672
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Command variables
The following command variables are generated in the method run by the
command MET Cond and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BP{x}.DME
Measured value differential for the break point ×
(1…9)
.BP{x}.MEA
Measured value for the break point × (1…9) in the
unit of the measured value
.BP{x}.TEM
Temperature for the break point × (1…9) in °C
.BP{x}.TIM
Time for the break point × (1…9) in s
.BP{x}.VOL
Volume for the break point × (1…9) in mL
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0
= READY; invalid (variable not available) = Com-
mand has never been started
.CLC
Cell constant of the sensor used in the command
for conductivity measurement cells
.CONC
Concentration of the solution used for the com-
mand
.CYL
Cylinder volume of the exchange or dosing unit
used for the command
.DBL
Total duration for the processing of the command
in s
.DSC
Time for processing all start conditions in s
.EME
End measured value (measured value after process-
ing of the command) in the unit of the measured
value
.EP{x}.DME
Measured value differential for the endpoint × (1…
9)
.EP{x}.MEA
Measured value for the endpoint × (1…9) in the
unit of the measured value
.EP{x}.MEP
Quantity of endpoints in the window × (1…9); 1 =
1 endpoint, 2 = 2 or more endpoints, 3 = EP
corrected with Autodrift, 4 = EP corrected
with manual drift
.EP{x}.TEM
Temperature for the endpoint × (1…9) in °C