6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1477
Rack offset
The rack offset is a production-dependent tolerance value between the
upper part of the rack and the lower part. The value is determined by a
rack adjustment and displayed here. If necessary, it can be edited.
Range
-5.00 ... 5.00 °
6.12.3.3
Properties - Lift positions
Tab: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack data ▶ Lift posi-
tions
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions
except those that are defined as Special beaker.
Work position
Work position for Lift 1. At this lift position the electrodes, stirrer and
buret tips are optimally positioned for work.
Range
0 ... 235 mm
Rinse position
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.
Range
0 ... 235 mm
Shift position
Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be
selected so that a safe movement across the entire rack is possible at any
time.
Range
0 ... 235 mm
Special position
Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solu-
tion.
Range
0 ... 235 mm
Tower 2
Defines the lift positions for Tower 2. These apply for all rack positions
except those that are defined as Special beaker.