6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1400
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Input
24 characters
Default value
'empty'
Serial number
Serial number of exchange unit. With intelligent exchange units the serial
number is read off automatically and cannot be edited.
Input
10 characters
Default value
'empty'
Cylinder volume
Cylinder volume of exchange unit. With intelligent exchange units the cyl-
inder volume is read off automatically and cannot be edited. If in the
method you have selected a solution in a titration or dosing command
then the cylinder volume will be checked in the run.
Selection
1 mL | 5 mL | 10 mL | 20 mL | 50 mL
Default value
20 mL
Cylinder serial number
Serial number of cylinder. It is printed on new cylinders. With intelligent
exchange units it is read off automatically. It can be modified at any time,
e.g. if the cylinder has to be replaced.
IEU
Input
8 characters
Default value
'empty'
EU
Input
10 characters
Default value
'empty'
Parameters for preparation
Configures the parameters to be used in the command PREP.
Volume
Volume to be dosed during preparation.
Range
0.00000 ... 99999.99999 mL
Selection
Cylinder volume
Default value
Cylinder volume
Cylinder volume
The whole contents of the cylinder will be dosed.
Cycles
Number of rinsing cycles during preparation.