Metrohm tiamo 2.3 (process analysis) User Manual
Page 1438
6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1422
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Next calibration
Date on which the next calibration is to be carried out. The date can be
selected by pressing
in the dialog window Select date (see Chapter
2.5.1, page 89). After a date has been entered, the value for the Calibra-
tion interval will be adapted automatically.
Can only be edited if monitoring is switched on.
Default value
Calibration date+ 999 days
Message
The message options can only be edited if monitoring is switched on.
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the e-mail address defined under [E-
mail] if this option is activated. The message is sent in text format.
[E-mail...]
[E-mail...] opens the window Send E-mail (see Chapter 2.6.1, page 92).
Acoustic signal
on | off (Default value: off)
If this option is activated, an acoustic signal will be emitted additionally to
the message defined above.
Action
The actions can only be edited if monitoring is switched on.
If during calibration monitoring it is found that its validity has expired then
one of the following actions will be triggered automatically at start test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the validity of the calibration has expired will be
automatically saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity of the calibration has expired will be saved automatically
with the determination.