2 cal meas ph - general/hardware – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1162
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1146
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Designa-
tion
Description
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
5.6.5.13.2
CAL MEAS pH - General/Hardware
Tab: Method
▶ CAL MEAS pH ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are
defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.