Metrohm tiamo 2.3 (process analysis) User Manual
Page 1191
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1175
Lower limit
Lower limit of the temperature. When the temperature falls below this
value, the action Lower temperature limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
-20.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok is triggered.
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
150.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
500.0 °C
Upper hysteresis
Upper hysteresis of the temperature. When the temperature once again
falls below the upper limit by this hysteresis value, the action Tempera-
ture upper limit ok is triggered
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Action
Selection of the event to be triggered if the upper or lower limit is excee-
ded: