6 met cond - conductometric evaluation, Conductometric evaluation – Metrohm tiamo 2.3 (process analysis) User Manual
Page 700
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
684
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Stop measured value
Stop when the entered measured value for a measuring point has been
exceeded or not reached since the start of the titration.
Range
0.00000 ... 2000.00
Selection
off
Default value
off
Stop time
Stop when the entered time has elapsed after the start of the titration
(including start conditions).
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Rate (volume/time unit) at which the buret is to be filled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.3.6.6
MET Cond - Conductometric evaluation
Tab: Method
▶ MET Cond ▶ Properties… ▶ Conductometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the conductometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of
the following evaluation methods can be chosen:
Conductometric evaluation
Selection
Evaluation without window | Evaluation with
volume window (mL)
Default value
Evaluation without window
Evaluation without window
With this option, the measurement curve is smoothed with the value
specified for the parameter Smoothing. Only the endpoint with the
highest value will be evaluated. If one wishes to evaluate several end-
points, then one must work with Evaluation with volume window
(mL).